Laboratory Reference Guide · Part 2

How to Read an HPLC Chromatogram on a Peptide COA

Published September 26, 2026 · ROVIQ Research

Most peptide COAs report purity as a single number, such as "99.8%", with a chromatogram printed underneath. Part 1 of this guide covered the COA as a whole: identity, purity, batch number and testing lab. It also recommended looking for the chromatogram and not just the percentage. This part explains what that chromatogram actually shows, how the purity figure is derived from it, and what it cannot tell you.

What the instrument is doing

High-performance liquid chromatography (HPLC) pushes a dissolved sample through a packed column. Components of the sample interact with the column packing to different degrees, so they leave it at different times. Reversed-phase HPLC is the most widely used mode for peptide separations. In it, more hydrophobic molecules are held longer, and most laboratories run it with acidic water/acetonitrile mobile phases containing trifluoroacetic acid (TFA) (Mant et al., 2007).

As each component leaves the column it passes a detector. For peptides this is usually a UV detector. Peptide bonds absorb strongly in the far ultraviolet, below 220 nm, so detection is generally set at 210–220 nm. Aromatic residues (tyrosine, phenylalanine, tryptophan) additionally absorb at 250–290 nm (Mant et al., 2007). A COA may name a single wavelength, or a photodiode-array (PDA) detector, which records many wavelengths at once.

Reading the axes

The chromatogram is the detector's signal plotted against time:

Schematic HPLC chromatogram: a solvent-front disturbance near the start, a small early impurity peak, a tall main peak with a small shoulder on its trailing side, and a small late-eluting peak. 05 1015 Retention time (min) Absorbance (mAU) 1 · solvent front 2 · related impurity 3 · main peak 4 · shoulder 5 · late eluter
Figure 1. A schematic chromatogram, drawn for illustration and not taken from any real sample. Numbered features are explained in the section below.

The features that matter

  1. Solvent front. A disturbance in the first minute or so, from the injection itself and anything that isn't retained by the column. It is normally excluded from the purity calculation. Very small, very polar peptides can elute close to it, where separation from other unretained material is weakest. That is one reason identity confirmation by mass spectrometry matters alongside HPLC.
  2. Related impurities. Small peaks near the main peak. In synthetic peptides these are typically structural relatives of the target. Synthesis-related impurities include deletion sequences (a missing amino acid), insertions (an extra one), diastereomers from racemization, and incompletely removed protecting groups. Degradation products include oxidized residues, dimers and cyclization products (D'Hondt et al., 2014). Because they resemble the target, they often elute close to it.
  3. Main peak. The compound the COA is about. It should be labeled, or its retention time given, so you can tell which peak the percentage refers to.
  4. Shoulders. A bump on the side of the main peak, rather than a separate peak, can be a closely related impurity that the method did not fully resolve. How the lab splits or includes that area changes the reported purity.
  5. Late eluters. More hydrophobic components come off later. A run that ends shortly after the main peak cannot show anything that would have eluted after it. So check that the time axis extends well past the main peak.

How the purity percentage is calculated

The usual figure is area percent: the area under the main peak divided by the total area of all integrated peaks, times 100. A COA reading "99.8% (HPLC)" is saying that 99.8% of the UV-absorbing material the method detected, and integrated, came out as the main peak.

What area percent is not: it is not the amount of peptide in the vial. Water, salts and counter-ions such as trifluoroacetate, which is often left over from synthesis and purification (D'Hondt et al., 2014), don't appear as integrated peaks, yet they are part of the powder's weight. A sample can be 99% pure by HPLC and still contain noticeably less than its nominal weight of peptide. That is what a separate net peptide content result measures, and it will be covered in a later part of this guide.

The same logic sets other limits on the number. Material that barely absorbs at the detection wavelength contributes little area. An impurity that co-elutes exactly under the main peak is counted as main peak. And different laboratories use different columns, gradients and integration settings, so a 99.1% from one lab and a 99.4% from another do not mean the second sample is purer. A second decimal place is a statement of arithmetic, not of accuracy.

A checklist for any chromatogram

Red flags

References

  1. Mant CT, Chen Y, Yan Z, et al. HPLC analysis and purification of peptides. Methods Mol Biol. 2007;386:3-55. PMID: 18604941
  2. D'Hondt M, Bracke N, Taevernier L, et al. Related impurities in peptide medicines. J Pharm Biomed Anal. 2014;101:2-30. PMID: 25044089
This article is for laboratory research reference only. All compounds discussed are intended strictly for research and laboratory use — not for human or animal consumption. Nothing here constitutes dosing, medical, or health guidance.
Related research and lab guides
View ROVIQ's Tested Compounds →